Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation of the power button. The device was connected to a cellular modem from MultiTech systems for use with LIFE NET services. The corrective action was completed by Velocitor Solutions in December 2010.
Source facts
Action
Velocitor initiated this field action and notified affected customers on March 11, 2010. This action was reported to Physio-Control as completed on 12/21/2010.
Initiated
2010-03-11
Root cause
Environmental control
Terminated
2016-12-16
Distribution
worldwide
Recall number
Z-0761-2017
Recall status
Terminated
CDRH record ID
151125
FDA record created
2016-12-14
Recall event number
75743