Skip to content

Public FDA record

Z-0761-2017: The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indo

Recall record Z-0761-2017 for Physio-Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0761-2017
Company in FDA record
Physio-Control, Inc.
FDA record creation date
Dec 14, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation of the power button. The device was connected to a cellular modem from MultiTech systems for use with LIFE NET services. The corrective action was completed by Velocitor Solutions in December 2010.

Source facts

Action
Velocitor initiated this field action and notified affected customers on March 11, 2010. This action was reported to Physio-Control as completed on 12/21/2010.
Initiated
2010-03-11
Root cause
Environmental control
Terminated
2016-12-16
Distribution
worldwide
Recall number
Z-0761-2017
Recall status
Terminated
CDRH record ID
151125
FDA record created
2016-12-14
Recall event number
75743

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗