Skip to content

Public FDA record

Z-0761-2020: i-STAT cTnI cartridge

Recall record Z-0761-2020 for Abbott Point Of Care Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0761-2020
Company in FDA record
Abbott Point Of Care Inc.
FDA record creation date
Dec 31, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

i-STAT cTnI blue cartridges may generate a higher than expected number of quality check codes. The observed rate of quality check codes could be in the 3 to 5% range.

Source facts

Action
On October 16, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail. It explains Abbott Point of Care has made the decision to transition U.S. customers using the i-STAT cTnI blue cartridge, list number 06P23-25, to the i-STAT cTnI white cartridge, list number 03P90-25, beginning in November 2019. The i-STAT cTnI blue cartridge, list number 06P23-25, will be discontinued for sale in the United States as of January 1, 2020. It further explains to customers while the i-STAT cTnI blue cartridges can continue to be used for patient testing, APOC recommends that all users of the i-STAT System be informed of the possibility of a higher than expected number of quality check codes while using the i-STAT cTnI blue cartridge and maintain a validated alternate backup method until the i-STAT cTnI white cartridges are received. Your Abbott Point of Care representative will contact you to discuss the details and timing of this transition for your facility. If your facility repeatedly observes the i-STAT cTnI blue cartridges from a single lot generating a higher than expected number of QCCs, please contact APOC Technical Services for support and to report the information. If you have forwarded any i-STAT cTnI blue cartridges to another facility, please provide a copy of this letter to them. ADDITIONAL INFORMATION If you have any questions regarding this information, please contact Abbott Point of Care Technical Support at 1-866-324-8114, or via email at apoc_cTnIproductinfo@abbott.com or contact your Abbott Point of Care support services representative. We appreciate your understanding in this matter and thank you for your continued support of Abbott and Abbott products.
Initiated
2019-10-16
Root cause
Device Design
Terminated
2020-05-04
Distribution
US: AK AL AZ CA CT DC FL GA HI IL IN KS KY MD MI MS NC NE NH NM NV NY OH OK OR PA SC SD TN TX VA WA WV WY
Recall number
Z-0761-2020
Recall status
Terminated
CDRH record ID
177454
FDA record created
2019-12-31
Recall event number
84166

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗