Z-0761-2021: BD Phoenix PMIC-107: IVD is intended for the in vitro rapid identification (ID) of gram-positive bacteri

Recall record Z-0761-2021 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0761-2021
Company in FDA record
Becton Dickinson & Co.
FDA record creation date
Dec 31, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection

Source facts

Action
BD issued letters Urgent Medical Device Correction mailed beginning December 8, 2020 with FedEx tracking. Letter states reason for recall, health risk and action to take: 1.Share this product correction notification with all BD Phoenix" users within your facility network and post a copy near all BD Phoenix" instruments until the revised User Manual is available. 2. Respond to the product correction notification electronically by using the link BD.com/IDS-21-3976 or by returning to the attached Customer Response Form to the BD contact noted on the form so that BD may acknowledge your receipt of this notification per FDA requirements. Report any adverse health consequences experienced with the use of this product to BD. Technical Support 800-638-8663 Monday Friday between 7:00am and 7:00pm (EST) in the United States. For customers outside of the United States, contact your local BD representative or distributor Actions Taken by BD: BD is updating the BD Phoenix" User Manual to provide additional guidance on use of the impacted panels and the manual is scheduled to be released by April 30, 2021. The revised manual will be accessible at bd.com/e-labeling. A BDXpert rule is also being developed to further assist customers on how to proceed with MIC readings for SXT. The BDXpert rule is set to be released by March 30, 2021 and will be included in the PUD update
Initiated
2020-12-08
Root cause
Nonconforming Material/Component
Terminated
2023-10-03
Distribution
Nationwide
Recall number
Z-0761-2021
Recall status
Terminated
CDRH record ID
184931
FDA record created
2020-12-31
Recall event number
86963

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗