Z-0762-2010: Stryker Reflex Hybrid Screw Extractor; Not Sterile: Manufactured by: Stryker Spine SAS, Cestas, France; D

Recall record Z-0762-2010 for Stryker Spine. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0762-2010
Company in FDA record
Stryker Spine
FDA classification date
Apr 12, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the screw.

Source facts

Action
An "Urgent Product Recall" letter was sent to Stryker Branches/Agencies on October 12, 2009 and to surgeons who have used the device on October 26, 2009. The recall letter described the affected products, issue and potential hazards. Consignees were instructed to examine their inventory and hospital locations for the affected products and return to Stryker Spine as indicated in the recall letter. Customers were also instructed to complete the enclosed Customer Response form and fax to 1-201-760-8370. Direct questions about the recall to Stryker Spine by calling 1-201-760-8206.
Initiated
2009-10-12
Root cause
Employee error
Terminated
2012-09-06
Distribution
Worldwide Distribution.
Recall number
Z-0762-2010
Recall status
Terminated
CDRH record ID
87752
FDA classified
2010-04-12
Recall event number
54169

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗