An internal real-time stability monitoring program revealed decreased recovery of myoglobin (-20% bias) for Randox Immunoturbidimetric and Siemens Centaur Classic methods and CKMB (-21% bias) for Roche e411, BioMerieux Vidas, and Siemens Centaur Classic methods.
Source facts
Action
Randox sent an Urgent Field Safety Notice dated September 2, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Please note the control value sheets have been updated with the new values. A FSN has been completed to advise customers on the appropriate action to take.
For further questions, please call (304) 728-2890.
Initiated
2016-09-16
Root cause
Nonconforming Material/Component
Terminated
2018-04-11
Distribution
Worldwide Distribution - US Distribution.
Recall number
Z-0762-2017
Recall status
Terminated
CDRH record ID
149843
FDA record created
2016-12-14
Recall event number
75262