There is a potential hardware error for the iMRX System with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible
Source facts
Action
An Urgent Field Safety Notice letter was sent to customers on 11/26/2019 via FedEx.. The letter state the issue, potential risk, the actions to be taken, and stated that the device will be modified. Questions can be directed to the service line at 1-866-475-0525.
Initiated
2019-11-26
Root cause
Component design/selection
Terminated
2020-08-04
Distribution
CT, MA, VA. International distribution: Australia, Canada, and Qatar.
Recall number
Z-0762-2020
Recall status
Terminated
CDRH record ID
178243
FDA record created
2019-12-31
Recall event number
84400