Open original FDA record ↗Recall record
- FDA identifier
- Z-0763-2010
- Company in FDA record
- Stryker Spine
- FDA classification date
- Apr 12, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the screw.
Source facts
- Action
- An "Urgent Product Recall" letter was sent to Stryker Branches/Agencies on October 12, 2009 and to surgeons who have used the device on October 26, 2009. The recall letter described the affected products, issue and potential hazards. Consignees were instructed to examine their inventory and hospital locations for the affected products and return to Stryker Spine as indicated in the recall letter. Customers were also instructed to complete the enclosed Customer Response form and fax to 1-201-760-8370. Direct questions about the recall to Stryker Spine by calling 1-201-760-8206.
- Initiated
- 2009-10-12
- Root cause
- Employee error
- Terminated
- 2012-09-06
- Distribution
- Worldwide Distribution.
- Recall number
- Z-0763-2010
- Recall status
- Terminated
- CDRH record ID
- 87753
- FDA classified
- 2010-04-12
- Recall event number
- 54169