Open original FDA record ↗Recall record
- FDA identifier
- Z-0764-2013
- Company in FDA record
- MOOG Medical Devices Group
- FDA classification date
- Jan 31, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm is adding an addendum to the operator's manual for Infinity Enteralite Ambulatory Feeding Pumps pertaining to the Infinite Dose mode and the inability of the Pump to detect air in line.
Source facts
- Action
- An Urgent Field Safety Notice was sent to all original consignees by certified mail beginning on Monday, October 15, 2012. Consignees were provided with an Addendum to the Operator's Manual as part of the Notice. Consignees were instructed to assure pump users are aware of the information in the addendum and to store the addendum with the Operator's Manual. Consignees were also requested to complete and return a Medical Device Recall reply form to the firm. For additional information or clarification, consignees are to contact Holly Imes, Manager, Regulatory Affairs at 801-264-1001, ext. 230 or via email at: himesstark@moog.com.
- Initiated
- 2012-10-12
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-04-10
- Distribution
- Worldwide Distribution and US Nationwide in the countries of (Germany, Sweeden, Croatia, Lithuania, and Dubai).
- Recall number
- Z-0764-2013
- Recall status
- Terminated
- CDRH record ID
- 113975
- FDA classified
- 2013-01-31
- Recall event number
- 63485