Z-0766-06: System 1000 family of Hemodialysis Instruments, including the System 1000, TINA, AURORA and Altratouch; B

Recall record Z-0766-06 for Baxter Healthcare Renal Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0766-06
Company in FDA record
Baxter Healthcare Renal Div
FDA classification date
Apr 18, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The older style 'Through Hardware' Dialysate Monitoring Manifolds and 'Noryl' Ultra Filtration Removal Regulators may become cracked due to mechanical and thermal stress. If either of these parts become cracked during patient treatment, there is the potential for an ultrafiltration error to occur, which can result in excessive removal of fluid from the patient.

Source facts

Action
Baxter sent Urgent Device Correction letters dated 02/24/06 to all System 1000, TINA, Altratouch and Aurora Hemodialysis Instrument customers via first class mail on the same date, to the attention of the Hemodialysis Administrator. The letters informed the accounts that the potential for the older style 'Through Hardware' Dialysate Monitoring Manifolds and 'Noryl' Ultra Filtration Removal Regulators to become cracked due to mechanical and thermal stress. Color photographs of the old style and new style components were included in the letter. The accounts were informed that if the cracking occurs during patient treatment, the potential exists for an ultra filtration error to occur, which can result in excessive removal of fluid from the patient. Baxter provided a chart to determine which upgrade kit is required for each unit, based on the serial number involved. The accounts were requested to complete the enclosed Action Completion Response form, indicating the instrument serial number and which parts are needed for each machine at the site, and fax the sheet to 1-727-544-3025. Upon receipt of the Action Completion Reply form, a Baxter Field Service Engineer will call the account to schedule the upgrades. Any technical questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 3, option 2, choice 1.
Initiated
2006-02-24
Root cause
Other
Terminated
2008-10-28
Distribution
Nationwide including Puerto Rico, and internationally to Algeria, Austria, Bahrain, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Ireland, Israel, Italy, Jordan, Lebanon, Morocco, the Netherlands, Nigeria, Oman, Poland, Romania, Russia, Slovakia, South Africa, Spain, Sweden, Switzerland, Tunisia, UAE, UK, Turkey, Colombia, Dominican Republic, Trinidad, Jamaica, St. Croix, Brazil, Peru, Ecuador, Australia, Argentina, Paraguay, Uruguay, China, Korea, Taiwan, Philippines, India, Singapore, Malaysia, Thailand, Indonesia and Canada.
Recall number
Z-0766-06
Recall status
Terminated
CDRH record ID
44321
FDA classified
2006-04-18
Recall event number
34645

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗