It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery.
Source facts
Action
DJO Global sent an Urgent Field Safety Notice dated September 19, 2016, to all affected customers. The field safety notice provided a newly revised Surgical Technique to educate users on proper use of the device. Customers were instructed to become familiar with the new surgical techniques and notify all of their applicable users of the revised methods. Customers were also instructed to sign the Acknowledgement and Receipt Form. Customers with questions were advised to call 760-734-3551.
For questions regarding this recall call 512-832-6302.
Initiated
2016-09-19
Root cause
Device Design
Terminated
2017-10-20
Distribution
Nationwide Distribution
Recall number
Z-0766-2017
Recall status
Terminated
CDRH record ID
150980
FDA record created
2016-12-14
Recall event number
75645
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.