Open original FDA record ↗Recall record
- FDA identifier
- Z-0766-2018
- Company in FDA record
- Steris Corporation
- FDA record creation date
- Feb 27, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Source facts
- Action
- Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.
- Initiated
- 2017-06-15
- Root cause
- Component change control
- Terminated
- 2018-07-17
- Distribution
- US Nationwide Distribution
- Recall number
- Z-0766-2018
- Recall status
- Terminated
- CDRH record ID
- 161365
- FDA record created
- 2018-02-27
- Recall event number
- 79100