- FDA identifier
- Z-0767-2013
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Feb 1, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.
Source facts
- Action
- Phillips issued an Urgent Field Safety Notice to all customers on January 14, 2013, to notify users that the Phillips EasyDiagnost Eleva systems, release 3.0 has a mirror icon that could be misinterpereted and could potentially lead to misdiagnosis. Philips will upgrade the system to version 3.1 which involves software and hardware upgrades of the system. This software provides a replacement icon that reduces the risk of the potential misinterpretation of the original mirror icon. Additionally, Philips will provide updated Instructions for Use (IFU) for the Eleva workspot and the EasyDiagnost 3.1 system and an addendum to the EasyDiagnost system IFU. Users will be contacted by Phillips to schedule the installation. For questions contact your local Philips representative: 1-800-722-9377.
- Initiated
- 2013-01-16
- Root cause
- Software design (manufacturing process)
- Terminated
- 2017-01-23
- Distribution
- US Nationwide Distribution including the states of FL, CA, WA, AZ, NC, KY, DE, AZ and MO.
- Recall number
- Z-0767-2013
- Recall status
- Terminated
- CDRH record ID
- 115393
- FDA classified
- 2013-02-01
- Recall event number
- 64127