The treatment table positioning belt can slip gear positions, resulting in deviation of patient setup and treatment positions from the time of slippage until the discovery of slippage and the halting of treatment.
Source facts
Action
Customer notification letter was sent via certified mail on November 18, 2015. The notification explains the issue and actions to be taken by the customer.
Initiated
2015-11-18
Root cause
Labeling design
Terminated
2017-04-03
Distribution
US nationwide distribution including FL, NJ, OK, and MO.
Recall number
Z-0767-2016
Recall status
Terminated
CDRH record ID
140206
FDA classified
2016-02-08
Recall event number
72210