Open original FDA record ↗Recall record
- FDA identifier
- Z-0767-2017
- Company in FDA record
- Firefly Medical, Inc.
- FDA classification date
- Dec 15, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The dowel pin could potentially slip out of its press fit, resulting in the product potentially slumping to one side which could cause the patient to stumble.
Source facts
- Action
- Firm notified customers on **11/18/16** via certified "Urgent: Medical Device Recall (Field Correction) IVEA Model 500A" letter to correct product in the field. Customers were asked to thoroughly read the letter and to immediately follow the actions in the enclosed Product Correction Kit. The instructions include steps for cleaning the correction area and installing the permanent adhesive caps, as well as, completing and returning the product correction form to Firefly Medical. Customers may continue to use their units until the Product Correction actions are complete, as long as pins are not protruding. If the pins are protruding, then customers are to stop using the device immediately and complete the product correction instructions. Customers can call 970-472-5323 or e-mail info@iveamobility.com if they have any questions.
- Initiated
- 2016-11-01
- Root cause
- Process control
- Terminated
- 2017-08-01
- Distribution
- Worldwide Distribution -- USA and Canada.
- Recall number
- Z-0767-2017
- Recall status
- Terminated
- CDRH record ID
- 150828
- FDA classified
- 2016-12-15
- Recall event number
- 75571