Z-0767-2020: O-arm" O2 Imaging System, Product Number/CFN: BI-700-02000

Recall record Z-0767-2020 for Medtronic Navigation, Inc.-Littleton. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0767-2020
Company in FDA record
Medtronic Navigation, Inc.-Littleton
FDA classification date
Dec 31, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing issue affecting particular O-arm" O2 Imaging Systems, Medtronic has determined that the gantry tractor motor drive belt may loosen over an extended number of uses at a higher frequency in these systems. If the belt loosens, it may result in rotation of 3D images about the gantry isocenter. While the 3D image is anatomically accurate within the image itself and may be used to confirm therapy, its electronic registered location may be rotated relative to the actual physical position of the patient.

Source facts

Action
On November 15, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail stating: MEDICAL DEVICE CORRECTION O-arm" O2 Imaging System Dear Healthcare Professional: Medtronic has determined through internal testing that there is a potential for navigational inaccuracy when utilizing the O-armTM O2 Imaging Systems auto-registration feature, when used in conjunction with an Image-Guided Surgery System for a specific set of O-armTM O2 Imaging System serial numbers. A list of potentially affected systems is provided in Table 1 of this letter. Due to loosening of an internal motor drive belt over extended number of uses, image rotation may occur relative to patient positioning data. Inaccuracy may occur when the O-arm registration information is transferred to an Image-Guided Surgery System to be used in navigation, utilizing the rotated images. Navigational inaccuracies are inherent in the use of any Image-Guided Surgery System utilizing an O-armTM O2 Imaging System, but the potential for inaccuracy may be higher if image rotation occurs in an affected system. This letter provides awareness of this issue and reinforces existing instructions within the O-arm" O2 Imaging System User Manual (IFU) that allow for identification of the issue if it occurs. This issue does not involve the O-arm 1000 Imaging Systems. Issue Background and Summary: The O-arm" O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. Medtronic has determined that if the gantry tractor motor drive belt loosens over extended number of uses, it can result in rotation of 3D images about the gantry isocenter. While the 3D image is anatomically accurate within the image itself and may be used to confirm therapy, its electronic registered location may
Initiated
2019-11-15
Root cause
Device Design
Terminated
2023-09-19
Distribution
US: AR CA CO CT FL ID IL IN KS KY MI MN MO NC ND NE NY OH OR PA SC SD TX VA WA WI WV OUS: AUSTRALIA CANADA CHINA FRANCE GERMANY INDIA ITALY JAPAN LEBANON NETHERLANDS ROMANIA SAUDI ARABIA SPAIN SWEDEN SWITZERLAND UNITED KINGDOM
Recall number
Z-0767-2020
Recall status
Terminated
CDRH record ID
178329
FDA classified
2019-12-31
Recall event number
84441

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗