Skip to content

Public FDA record

Z-0767-2021: Eleganza 5 (AC powered adjustable bed), Item Numbers: a) 1GE512055-40, b) 1GE512055-45, c) 1GE512055-85,

Recall record Z-0767-2021 for Linet Spol. S.r.o.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0767-2021
Company in FDA record
Linet Spol. S.r.o.
FDA record creation date
Jan 4, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A component used within the side rail mechanism may malfunction, resulting in the side rail either not locking properly in the upper position or the side rail becoming stuck in the upper position.

Source facts

Action
The firm initiated the recall by email and hand delivery on 11/13/2019. The notice explains the malfunction and risk, and provides the following information: "Advice on action to be taken by the user: Linet has developed a correction for the side rail mechanism. We will contact you to schedule a Linet or partner technician to update your affected Eleganza 5 Beds with this correction. Until your affected beds are updated, we ask that you follow the safety information as outlined in the Eleganza 5 User manual (D9U001GE5-0101, version: 11, Publication date: 2019-11 (available at your local distributor)) when operating the side rails. Section 11.1 Side rails Chapter MANIPULATION To raise siderails up: >Grab side rail by Side Rail Handle, >Pull side rail up until it latches. You will hear audible "click". If you raise the side rails to the upper position and you find the side rail is not secured and/or no audible click is heard during this operation, take the bed out of service until the bed has been updated." Contact Linet Technical Services at phone no.: 877-815-9897
Initiated
2019-11-13
Root cause
Device Design
Distribution
US Nationwide distribution.
Recall number
Z-0767-2021
Recall status
Open, Classified
CDRH record ID
184252
FDA record created
2021-01-04
Recall event number
86647

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗