Skip to content

Public FDA record

Z-0768-2019: Vial2BAG Transfer Device, REF 6070104, Single use only, Rx Only

Recall record Z-0768-2019 for WEST PHARMA. SERVICES IL, LTD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0768-2019
Company in FDA record
WEST PHARMA. SERVICES IL, LTD
FDA classification date
Jan 28, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Adverse events have been reported with the use of oxytocin in conjunction with the 13mm Vial2Bag DC in pregnant women. Vial2Bag is recalled due to similarities with 13mm Vial2Bag DC.

Source facts

Action
On December 26, 2018, West Pharmaceutical Services, Inc. issued URGENT MEDICAL DEVICE CORRECTION and ACKNOWLEDGEMENT FORMS to their customers asking customers to avoid use of the drug oxytocin with the device. Actions to be taken by the Distributor: 1. Complete the required Acknowledgement and Receipt Form below and return (via email) to Kevin.Lentz@westpharma.com when complete (See attached form). 2. Prior to filling order / releasing shipments of any Vial2Bag DC 13mm products (which have been placed on hold per our request) to hospital customers, affix a copy of this Voluntary Field Corrective Action Notice to each carton of Vial2Bag DC 13mm products in your possession. 3. Notify and provide the letter to all accounts/customers where the Affected Product was distributed. 4. Because the assessment is ongoing, please be prepared for updates to the notification. Actions to be taken by the Customer/User: 1. As a temporary measure you should avoid use of the drug oxytocin with the Vial2Bag DC 13mm until further notice until a full assessment has been completed. 2. Affix the notice to each carton of Vial2Bag DC 13mm products in your possession. 3. In addition, provide the notice to personnel with responsibilities related to product selection and administration, and according to your internal procedures for notification of field corrective actions. 4. Please complete the Recall Acknowledgement and Receipt Form and return to Karen Crimi by email at cs@progressivemedinc.com when complete responses are available. On January 24, 2019, West Pharmaceutical Services, Inc. initiated a product removal and broadened their action by distributing AMENDED URGENT MEDICAL DEVICE RECALL notices to their customers to include removal of three (3) devices from the field. On February 1, 2019, West Pharmaceutical Services, Inc. expanded their recall to include an additional sixteen (16) lots and issued notices titled - 2nd AMENDED URGENT MEDICAL DEVICE RECALL not
Initiated
2018-12-26
Root cause
Device Design
Terminated
2022-07-20
Distribution
US Distributor in MO
Recall number
Z-0768-2019
Recall status
Terminated
CDRH record ID
170467
FDA classified
2019-01-28
Recall event number
81877

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗