Reason stated in the source
Customer complaints of a leak between the blood and water compartments of the heat exchanger used in the oxygenators due to prolonged contact of the heat exchangers with sodium chloride causing micro-holes.
Source facts
- Action
- Urgent Field Safety Notices dated December 2, 2011 were sent to customers. The letter identified the affected products and described the problem. The letter also provided additional directions to the Instructions for Use in order to minimize risk and clarify priming procedures. Customers were to communicate the new warning to all personnel in the use of the affected products and complete the Customer Response Form. The form should be returned via fax or e-mail. Questions should be directed towards Sorin Group Customer Service at 1-800-650-2623.
- Initiated
- 2011-12-02
- Root cause
- Device Design
- Terminated
- 2012-06-13
- Distribution
- Worldwide Distribution
- Recall number
- Z-0770-2012
- Recall status
- Terminated
- CDRH record ID
- 106363
- FDA classified
- 2012-01-14
- Recall event number
- 60737