Z-0770-2019: Perfusion Pack, Rx only, Sterile EO

Recall record Z-0770-2019 for LivaNova USA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0770-2019
Company in FDA record
LivaNova USA
FDA record creation date
Jan 23, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

4-way stopcocks and stopcock manifold assembled in certain perfusion tubing systems, indicated for use in surgical procedures requiring extracorporeal support for periods of up to six hours, may leak or be occluded.

Source facts

Action
On 11/09/2018 Medical Device Correction Notices were mailed to customers via UPS First Overnight. Customers were advised: 1) Inspect your stock and identify Perfusion Tubing System(s) belonging to the impacted lots in inventory, as noted in Attachment 1. 2) The recalling firm will be shipping one case of sterile stopcocks (part number: SCO4S) and manifolds (part number: 020581600) to your facility as an immediate replacement for those remaining that may be defective in your impacted Perfusion Tubing System stock. 3) As each affected pack is opened for use, please replace the original stopcocks in the pack with the spare provided by the firm. The original stopcocks should be destroyed/ discarded. 4) Contact your firm Representative if additional replacement stopcocks are required. In addition, customers were asked to complete and return the attached Customer Response Form. Also, to assure within their organization that the correction notice is communicated to all personnel who need to be aware of it. For questions about this Medical Device Correction, please contact (800) 986-4702 or e-mail USFSN@livanova.com
Initiated
2018-11-09
Root cause
Process control
Terminated
2022-05-03
Distribution
Worldwide Distribution - Domestic (US) Consignees: AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV, MS; Foreign (OUS) Consignees: Canada
Recall number
Z-0770-2019
Recall status
Terminated
CDRH record ID
169426
FDA record created
2019-01-23
Recall event number
81667

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗