Open original FDA record ↗Recall record
- FDA identifier
- Z-0770-2020
- Company in FDA record
- Carl Zeiss Meditec, Inc.
- FDA record creation date
- Jan 2, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch.
Source facts
- Action
- Recalling Firm issued Urgent Medical Device Recall notices to all customers dated October 16, 2018. Customers were advised to take the following actions: Until the mandatory hardware update has been performed, follow the procedure below to continue the use of the PLEX Elite 9000: When adjusting the height of the Power Table, touch only the black plastic portion of the Up/Down switch. Avoid any contact with the entire metal enclosure (including mounting screws) surrounding the black plastic Up/Down switch. Failure to follow this procedure may result in potential injury. CAUTION: Operator(s) with an electrical implant are more sensitive to electrostatic discharge and must avoid use of the system prior to the hardware update. To correct the issue, customers will be contacted by a representative to schedule a Field Service Engineer to perform a mandatory hardware update for their PLEX Elite 9000.
- Initiated
- 2018-10-16
- Root cause
- Device Design
- Terminated
- 2021-06-11
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, TX, NY, MA, OH, PA, IA, , KY, FL, MI, IL, MD and Germany, Japan, Switzerland, France Italy, Australia, Austria, Korea, China, India, Canada, Singapore, Norway, Taiwan, Israel, UAE.
- Recall number
- Z-0770-2020
- Recall status
- Terminated
- CDRH record ID
- 176021
- FDA record created
- 2020-01-02
- Recall event number
- 83229