Open original FDA record ↗Recall record
- FDA identifier
- Z-0771-2007
- Company in FDA record
- Toshiba American Med Sys Inc
- FDA classification date
- May 5, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Transducer Surface Temperature : It was found during testing that when using the system in the 2DCDI + PW mode, with the Scale Graph function active, that the transducer surface temperature could exceed the 27 degrees Celsius limit as specified in the IEC safety standard.
Source facts
- Action
- Notification letters were sent to customers on December 19, 2206. Customers were instructed to avoid using Scale Graph function with live images, and instructed to call up the Scale Graph after images are frozen. Customers were informed that a Toshiba Customer Service Engineer will contact them to schedule a software upgrade when software becomes available. If customers have any questions they were instructed to contact their Toshiba Customer Service Engineer or call (800) 421-1968 and ask for the Regulatory Affairs Department. Customers were also instructed to distribute this information to all users of the product and also with the Bio-medical Department at their facility.
- Initiated
- 2006-12-19
- Root cause
- Other
- Terminated
- 2012-02-15
- Distribution
- Nationwide Distribution including the states of CA, FL, GA, ID, MA, MN, NC, NY, PA, RI, TN, TX, WA, and WI.
- Recall number
- Z-0771-2007
- Recall status
- Terminated
- CDRH record ID
- 51592
- FDA classified
- 2007-05-05
- Recall event number
- 37813