Open original FDA record ↗Recall record
- FDA identifier
- Z-0771-2016
- Company in FDA record
- Accuray Incorporated
- FDA classification date
- Feb 9, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The gun box mounting bracket may fail to support the weight of the gun box when in vertical (inverted) position. If this failure occurs the gun box may become loose and could come into contact with a patient.
Source facts
- Action
- Accuray sent an Urgent Safety Advisory Notification letters dated January 22, 2016 to all affected customers by trackable mail method. The letters identified the affected product, problem and actions to be taken. Customers were instructed to complete and return the Acknowledgement form. For questions contact Customer Support at 1-877-668-8667 or visit the Accuray Technical Solution Center at http://www.accuray.com/services-support
- Initiated
- 2016-01-22
- Root cause
- Device Design
- Terminated
- 2016-10-20
- Distribution
- Worldwide Distribution. US Nationwide, Venezuela, United Kingdom, Ukraine, Turkey, Thailand, Taiwan, Switzerland, Spain, Saudi Arabia, Russia, Poland, Pakistan, Myanmar, Mexico, Korea, Japan, Italy, Ireland, India, Greece, Germany, France, Finland, Czech Republic, Colombia, China, Canada, and Belgium.
- Recall number
- Z-0771-2016
- Recall status
- Terminated
- CDRH record ID
- 143569
- FDA classified
- 2016-02-09
- Recall event number
- 73193