Open original FDA record ↗Recall record
- FDA identifier
- Z-0771-2017
- Company in FDA record
- Teleflex Medical
- FDA record creation date
- Dec 16, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Labeling Inconsistency: The products have been labeled with the incorrect expiration date which exceeds the actual, validated 35-month shelf life.
Source facts
- Action
- Teleflex sent an Urgent Medical Device Recall Notification letter dated November 22, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers with affected stock, immediately discontinue use and quarantine any affected products. To return product, customers were advised to complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recall@teleflex.com. Customers with no affected stock should also complete the enclosed Recall Acknowledgment Form to confirm receipt of the letter. Customers with questions should contact Customer Service at 1-866-246-6990.
- Initiated
- 2016-11-22
- Root cause
- Error in labeling
- Terminated
- 2017-11-24
- Distribution
- Nationwide Distribution to CA, CO, FL, GA, IL, KS, KY, MA, MO, MS, NC, NJ, NY, OH, OK, OR, PA and TX.
- Recall number
- Z-0771-2017
- Recall status
- Terminated
- CDRH record ID
- 151249
- FDA record created
- 2016-12-16
- Recall event number
- 75781