Open original FDA record ↗Recall record
- FDA identifier
- Z-0771-2019
- Company in FDA record
- Varian Medical Systems, Inc.
- FDA record creation date
- Jan 24, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo- rithm calculates the water equivalent range incorrectly for non-square 3D CT images (either different number of pixels in X and Y, or non-square pixels). Before the dose is calculated, Eclipse resamples the CT images to create a calculation image with a maximum resolution of 256 x 256. The PCS algorithm assumes uniform resolution in X and Y directions, or that X = Y for all images, and erroneously sets the Y = X for dose calcu- lation. For images that are not square, where either X<>Y, or the length of X does not equal the length of Y, the computed water equivalent range R is erroneous compared to the correct range R: 1) X=Y; R =R (R is correct); 2) XY R >R (R is too big).
Source facts
- Action
- The firm, Varian, disseminated the "URGENT MEDICAL DEVICE CORRECTION FIELD SAFETY NOTICE" dated 12/17/2018 to its customers on 12/26/2018. The letter explained the product, problem and a work around. The customers were instructed to do the following: Please retain a copy of this document along with your most current product labeling. complete the attached Recall Return Response form provided. Kindly return the completed form to returnresponse@varian.com Advise the appropriate personnel working in your radiotherapy department of the content of this letter. For future reference, this document will be posted to the Varian Medical Systems customer support website: http://www.MyVarian.com. Varian Medical Systems has developed a technical correction for this issue. You will be contacted by a Customer Service representative to schedule its installation on your system. If you require further clarification, please feel free to contact your local Varian Medical Systems Customer Support District or Regional Manager or Varian Oncology Help Desk Contact Information at +1.888.827.4265 or Email: support @varian.com.
- Initiated
- 2018-12-26
- Root cause
- Software design
- Terminated
- 2021-03-10
- Distribution
- Worldwide distribution: US (nationwide) and countries of: Australia, Brazil, Denmark France, Germany, Japan, Poland, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Sweden, Switzerland, Taiwan, and United Kingdom.
- Recall number
- Z-0771-2019
- Recall status
- Terminated
- CDRH record ID
- 170287
- FDA record created
- 2019-01-24
- Recall event number
- 81912