Z-0771-2019: Eclipse(TM) Treatment Planning System with Proton Convolution Superposition algorithm license only The E

Recall record Z-0771-2019 for Varian Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0771-2019
Company in FDA record
Varian Medical Systems, Inc.
FDA record creation date
Jan 24, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo- rithm calculates the water equivalent range incorrectly for non-square 3D CT images (either different number of pixels in X and Y, or non-square pixels). Before the dose is calculated, Eclipse resamples the CT images to create a calculation image with a maximum resolution of 256 x 256. The PCS algorithm assumes uniform resolution in X and Y directions, or that X = Y for all images, and erroneously sets the Y = X for dose calcu- lation. For images that are not square, where either X<>Y, or the length of X does not equal the length of Y, the computed water equivalent range R is erroneous compared to the correct range R: 1) X=Y; R =R (R is correct); 2) XY R >R (R is too big).

Source facts

Action
The firm, Varian, disseminated the "URGENT MEDICAL DEVICE CORRECTION FIELD SAFETY NOTICE" dated 12/17/2018 to its customers on 12/26/2018. The letter explained the product, problem and a work around. The customers were instructed to do the following: Please retain a copy of this document along with your most current product labeling. complete the attached Recall Return Response form provided. Kindly return the completed form to returnresponse@varian.com Advise the appropriate personnel working in your radiotherapy department of the content of this letter. For future reference, this document will be posted to the Varian Medical Systems customer support website: http://www.MyVarian.com. Varian Medical Systems has developed a technical correction for this issue. You will be contacted by a Customer Service representative to schedule its installation on your system. If you require further clarification, please feel free to contact your local Varian Medical Systems Customer Support District or Regional Manager or Varian Oncology Help Desk Contact Information at +1.888.827.4265 or Email: support @varian.com.
Initiated
2018-12-26
Root cause
Software design
Terminated
2021-03-10
Distribution
Worldwide distribution: US (nationwide) and countries of: Australia, Brazil, Denmark France, Germany, Japan, Poland, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Sweden, Switzerland, Taiwan, and United Kingdom.
Recall number
Z-0771-2019
Recall status
Terminated
CDRH record ID
170287
FDA record created
2019-01-24
Recall event number
81912

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗