The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)
Source facts
Action
The recalling firm issued letters dated 2/7/2019 via FedEx explaining the issue, informing the consignee there was no risk associated with use of the device, and that the firm is preparing for the replacement parts in order to schedule for the correction to be made.
Initiated
2019-02-25
Root cause
Device Design
Terminated
2021-06-15
Distribution
US Nationwide distribution in the states of AZ, CA, IL, KS, MD, NJ, NY, PA, SC, and VA, as well as PR.
Recall number
Z-0771-2020
Recall status
Terminated
CDRH record ID
177217
FDA record created
2020-01-02
Recall event number
84112