Z-0772-2012: Dimension(R) Creatinine Flex(R) Reagent Cartridges. The CREA method is an in vitro diagnostic test for t

Recall record Z-0772-2012 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0772-2012
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Jan 15, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Siemens has confirmed the potential for falsely elevated creatinine results with CREA FLex(R) reagent cartridge lots GA2196, GA2203, DC2221, DB2221, GA2229 and FA2237 when processing tests from a well that has been punctured.

Source facts

Action
Siemens sent an "URGENT FIELD SAFETY NOTICE" dated November 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter recommends that customers discontinue and discard any remaining inventory of the affected product. The firm also recommends customers to test QC at least every 8 hours with lots manufactured after lot FA2237. Additionally, a Field Correction Effectiveness Check form was attached to the letter for customers to complete and return via fax to 302-631-8467. Contact the Siemens Customer Service Center at 888-588-3916 for questions regarding this notice.
Initiated
2011-11-22
Root cause
Under Investigation by firm
Terminated
2014-04-03
Distribution
Worldwide Distribution-USA (nationwide) including the District of Columbia and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY and countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Benin, Bosnia Herzegovina, Botswana, Brazil, Canada, Canary Islands, Ceuta/Mililla, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, France, French Guinea, Germany, Great Britian, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxemburg, Macedonia, Mexico, Montenegro, Morocco, Nepal, Nether Antilles, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam, Yemen and Zimbabwe.
Recall number
Z-0772-2012
Recall status
Terminated
CDRH record ID
105994
FDA classified
2012-01-15
Recall event number
60618

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗