Open original FDA record ↗Recall record
- FDA identifier
- Z-0772-2018
- Company in FDA record
- Steris Corporation
- FDA classification date
- Feb 13, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Correction to update sterilizer software to prohibit the use of expired sterilant prior to the start of a sterilization cycle.
Source facts
- Action
- The V-PRO 60 system software has been updated to ensure the VAPROX Sterilant expiration date is verified prior to the initiation of each sterilization cycle. STERIS Service Technicians will be contacting your facility to update your V-PRO 60 Sterilizer(s) with the revised system software. User Action Users should continue to review their V-PRO 60 cycle printouts upon completion of each cycle to verify the VAPROX Sterilant was not expired. The VAPROX Sterilant expiration date is identified near the top of each cycle printout following all sterilization cycles as notated in the following illustration:
- Initiated
- 2017-09-28
- Root cause
- Device Design
- Terminated
- 2018-09-12
- Distribution
- Distributed in 44 states: AL, AK, AZ, AR, CA, CO, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY; Puerto Rico, and the following countries: Algeria, Australia, Azerbaijan, Bahrain, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Hong Kong, Hungary, India, Israel, Italy, Republic of Korea, Kuwait, Latvia, Malaysia, Mexico, Morocco, Myanmar, Panama, Peru, Philippines, Poland, Portugal, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, Thailand, Turkey, United Arab Emirates, and Vietnam.
- Recall number
- Z-0772-2018
- Recall status
- Terminated
- CDRH record ID
- 161372
- FDA classified
- 2018-02-13
- Recall event number
- 79101