Open original FDA record ↗Recall record
- FDA identifier
- Z-0772-2019
- Company in FDA record
- Leica Microsystems, Inc.
- FDA record creation date
- Jan 24, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The M220 optics may unintentionally drop into the surgical field, risking contact with the patient.
Source facts
- Action
- The firm, Leica, disseminated an "URGENT FIELD SAFETY NOTICE" dated 12/18/2018 on 01/02/2019 to its customers. The notice described the product, problem and actions to be taken. The notice directed the user to inspect the device for a gap between the black adapter and the white housing or a free rotation of the optics carrier for more 360o. If either are noted, the device could not be used. If you have any questions, contact the Clinical & Scientific Affairs Specialist at 847-876-5705 or email to: Jason.Battle@leica-miosystems.com.
- Initiated
- 2019-01-02
- Root cause
- Process control
- Terminated
- 2021-04-02
- Distribution
- Worldwide distribution: US (nationwide) and countries of: Albania, Algeria, Armenia, Argentina, Bangladesh, Brazil, Burundi, Canada, China, Colombia, Egypt, France, Germany, Guatemala, India, Indonesia, Japan, Nepal, Netherlands, New Zealand, Malaysia, Pakistan, Peru, Philippines, Poland, Qatar, Romania, Russia Federation, Saudi Arabia, Senegal, Spain, South Africa, South Korea, Switzerland, Tanzania, Thailand, Turkey, Ukraine, and Vietnam.
- Recall number
- Z-0772-2019
- Recall status
- Terminated
- CDRH record ID
- 170260
- FDA record created
- 2019-01-24
- Recall event number
- 81903