Z-0772-2024: VITROS Immunodiagnostic Products Anti-HBs Quantitative Reagent Pack, Catalog No. 6801925 VITROS Immunodi

Recall record Z-0772-2024 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0772-2024
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Jan 19, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected lots may experience increased calibration failures or an increase in falsely elevated results.

Source facts

Action
On 27 November 2023, consignees who had been shipped the affected VITROS Anti-HBs Reagent Pack and Calibrator were notified of the issue via Urgent Production Correction Notification letters. Customers were instructed to immediately discontinue using, render unusable, and discard all affected lots of the affected product from their inventory. - QuidelOrtho will provide credit or replacement for the discarded product. - QuidelOrtho recommends a review of previous results between 12-30 mIU/mL, which were generated using an affected lot of VITROS anti-HBs Reagent Pack, where only anti-HBs testing was performed and was positive (for example, postvaccination testing) or where only anti-HBs was positive in the triple-panel test. If you have further questions, please contact QuidelOrtho's Global Services Organization (formerly Ortho Care) at 1-800-421-3311.
Initiated
2023-11-27
Root cause
Nonconforming Material/Component
Distribution
Domestic distribution nationwide. OUS distribution to Bermuda, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Mexico, Singapore, UK, France, Germany, Italy, Spain, Portugal, Russia, Denmark, Norway, Sweden, France, Belgium, and The Netherlands.
Recall number
Z-0772-2024
Recall status
Open, Classified
CDRH record ID
204993
FDA classified
2024-01-19
Recall event number
93632

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗