Open original FDA record ↗Recall record
- FDA identifier
- Z-0773-2012
- Company in FDA record
- Microgenics Corp
- FDA classification date
- Jan 17, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Reports that third party control material for 4 lots of Tobramycin Calibrators that are recovering low and outside of the published range.
Source facts
- Action
- ThermoFisher Scientific sent a product removal letter dated November 11, 2011. Letters were sent to the sole distributor of the product. The letter identified the product the problem and the action needed to be taken by the customer. If you are using or have inventory of the affected lot numbers immediately discontinue use and destroy any remaining affected lot numbers per your local waste ordinances. If you have any further questions regarding this information please call 1-877-4Abbott. Customers outside the US should contact their local customer service.
- Initiated
- 2011-11-11
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-05-09
- Distribution
- Worldwide Distribution -- Nationwide Distribution including the country Europe.
- Recall number
- Z-0773-2012
- Recall status
- Terminated
- CDRH record ID
- 105597
- FDA classified
- 2012-01-17
- Recall event number
- 60503