Skip to content

Public FDA record

Z-0773-2012: Tobramycin Calibrators, REF 7F93-01; Microgenics Corporation, 46360 Fremont Blvd Fremont, CA 94538; Dis

Recall record Z-0773-2012 for Microgenics Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0773-2012
Company in FDA record
Microgenics Corp
FDA classification date
Jan 17, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Reports that third party control material for 4 lots of Tobramycin Calibrators that are recovering low and outside of the published range.

Source facts

Action
ThermoFisher Scientific sent a product removal letter dated November 11, 2011. Letters were sent to the sole distributor of the product. The letter identified the product the problem and the action needed to be taken by the customer. If you are using or have inventory of the affected lot numbers immediately discontinue use and destroy any remaining affected lot numbers per your local waste ordinances. If you have any further questions regarding this information please call 1-877-4Abbott. Customers outside the US should contact their local customer service.
Initiated
2011-11-11
Root cause
Nonconforming Material/Component
Terminated
2012-05-09
Distribution
Worldwide Distribution -- Nationwide Distribution including the country Europe.
Recall number
Z-0773-2012
Recall status
Terminated
CDRH record ID
105597
FDA classified
2012-01-17
Recall event number
60503

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗