Open original FDA record ↗Recall record
- FDA identifier
- Z-0773-2014
- Company in FDA record
- Ge Healthcare It
- FDA classification date
- Jan 15, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In Imagecast PACS versions prior to 3.6.51.01, the estimated magnification factor (0018,1114) was not accounted for. These versions of Imagecast PACS look for values to be present in the following tags to calculate a measurement from pixels to millimeters (mm): (0020,0030) Pixel Spacing and, (0018,1164) Imager Pixel Spacing. Because the estimated magnification factor is not used, distance meas
Source facts
- Action
- Important Product Information Letters dated March 4, 2011 were sent to all customers that received the defective software. The letters advised customers to upgrade their PACS software version to Imagecast 3.6. The software upgrade can be requested from GE Support via the eService portal or by telephone at 1-800-437-1171. If the customer elects to not upgrade the software, instructions were provided for a new system calibration procedure.
- Initiated
- 2011-03-04
- Root cause
- Software design
- Terminated
- 2016-05-16
- Distribution
- Nationwide Distribution: US including states of: AL, AZ, CT, FL, GA, IA, IN, LA, MA, MI, MN, MO, NC, NH, NJ, NY, OH, PA, TX, VA, WI, WV + District of Columbia.
- Recall number
- Z-0773-2014
- Recall status
- Terminated
- CDRH record ID
- 116903
- FDA classified
- 2014-01-15
- Recall event number
- 64740