Open original FDA record ↗Recall record
- FDA identifier
- Z-0773-2019
- Company in FDA record
- NEUSOFT MEDICAL SYSTEMS IMP & EX
- FDA record creation date
- Feb 5, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
For surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV the actual start and end position may not be consistent with the planned positons.
Source facts
- Action
- The firm, Neurosoft Medical Systems Co., Inc., issued a notice to its customers dated 11/26/2018. The notice described the product, problem and actions to be taken. The customers were instructed to do the following: Please plan surview length accurately according the clinical requirement, and not interrupt(abort or skip) scanning during the surview scan if the planning length is over 500mm. The defect will be remedied by updating software. Field Change Order related to software updated will be released to the affected systems in Jan, 2019. The affected systems will be upgraded with updated software in the field free of charge in 6 months after the FCO release. If you have any questions, contact Senior TQM Director at Tel: +86 024-2335831 1 or email: wangzq@neusoft.com.
- Initiated
- 2019-01-09
- Root cause
- Software design
- Terminated
- 2022-08-25
- Distribution
- US Distribution in states of: KS, OK and MA.
- Recall number
- Z-0773-2019
- Recall status
- Terminated
- CDRH record ID
- 170298
- FDA record created
- 2019-02-05
- Recall event number
- 81920