Skip to content

Public FDA record

Z-0773-2019: NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Sl

Recall record Z-0773-2019 for NEUSOFT MEDICAL SYSTEMS IMP & EX. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0773-2019
Company in FDA record
NEUSOFT MEDICAL SYSTEMS IMP & EX
FDA record creation date
Feb 5, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

For surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV the actual start and end position may not be consistent with the planned positons.

Source facts

Action
The firm, Neurosoft Medical Systems Co., Inc., issued a notice to its customers dated 11/26/2018. The notice described the product, problem and actions to be taken. The customers were instructed to do the following: Please plan surview length accurately according the clinical requirement, and not interrupt(abort or skip) scanning during the surview scan if the planning length is over 500mm. The defect will be remedied by updating software. Field Change Order related to software updated will be released to the affected systems in Jan, 2019. The affected systems will be upgraded with updated software in the field free of charge in 6 months after the FCO release. If you have any questions, contact Senior TQM Director at Tel: +86 024-2335831 1 or email: wangzq@neusoft.com.
Initiated
2019-01-09
Root cause
Software design
Terminated
2022-08-25
Distribution
US Distribution in states of: KS, OK and MA.
Recall number
Z-0773-2019
Recall status
Terminated
CDRH record ID
170298
FDA record created
2019-02-05
Recall event number
81920

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗