Skip to content

Public FDA record

Z-0775-2012: CADD High Volume Administration Set, REF 21-7081V-01, Sterile EO, Rx only, Made in Mexico, Smith Medical

Recall record Z-0775-2012 for Smiths Medical ASD, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0775-2012
Company in FDA record
Smiths Medical ASD, Inc.
FDA classification date
Jan 17, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Smith Medical has become aware of a small number of customer reports regarding occlusions/high pressure alarms with the use of one lot of CADD High Volume Administration Sets. EXPANDED: Smiths Medical became aware that an additional Product Reorder No. 21-738-01 and Lot No. 21X221 are also being recalled for receiving reports regarding occlusion/high pressure alarms with the use of these set

Source facts

Action
Consignees were sent a Smiths Medical "Urgent Field Safety Notice" letter dated January 04, 2011. The letter was addressed to Risk/ Safety Managers, Clinicians, Nursing, Pharmacy, Pain Management and Anesthesia Professionals, Distributors and other users of these devices. The letter described the details on affected devices, Advise on Action to be Taken by the User and distributor and Transmission of this Urgent Field Safety Notice. Requested consignees to inspect their inventory and remove affected product and to complete and return the attached Confirmation Form. EXPANDED: The additional devices were distributed 12/14/2011 thru 1/27/2012. An Urgent Medical Device letter, dated 3/16/2012, was sent to customers beginning 3/16/2012. The letter identified the additional affected product, described what the issue is and asked that customers return all unused sets to Smiths Medical for credit or replacement. Also, the customer was asked to complete and return an Urgent Medical Device Recall Confirmation Form. If the customer is a distributor, they were asked to pass along the information to their customers.
Initiated
2011-12-09
Root cause
Mixed-up of materials/components
Terminated
2012-11-06
Distribution
Nationwide Distribution -- AZ, CA, FL, GA, IL, KS, LA, ME, MA, NC, OR, PA,SC, TX, VA, WA, and WI. EXPANDED to include AR, CO, IA, IN, MD, MI, MN, MO, NJ, OH, OK, TN, WV and Canada.
Recall number
Z-0775-2012
Recall status
Terminated
CDRH record ID
106706
FDA classified
2012-01-17
Recall event number
60883

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗