Z-0775-2014: Estrogen Receptor AB-11 (Clone ID5) Mouse Monoclonal Antibody 1 mk (2mg.ml) Manufactured by: Lab Vision

Recall record Z-0775-2014 for Lab Vision Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0775-2014
Company in FDA record
Lab Vision Corporation
FDA classification date
Jan 16, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports that the product is not performing as expected. In house testing showed no staining on breast cancer tissues.

Source facts

Action
Lab Vision Corporation part of Thermo Fisher Scientific sent an Urgent Medical Device Recall letter dated December 6, 2013, to all affected customers. Revised letters will be sent on January 8, 2014. Changes include that the reagent is an Analyte Specific Reagent (ASR) and not an in vitro diagnostic device in the US. They also clarify that a false positive was reported by a customer and a false negative was confirmed during in house testing. Letters also correct the expiry date from 12/2013 to 03/2015. Data sheet was also updated to eliminate information stating that the presence of ER in breast tumors indicates an increased likelihood of response to estrogen therapy and also that the positive control must be determined by the investigator. If you have any questions, please contact Technical Service at 1-800-522-7270, option 2, option 7.
Initiated
2013-12-06
Root cause
Under Investigation by firm
Terminated
2014-02-25
Distribution
Worldwide Distribution: US (nationwide) including the states of TX and MA and internationally to: Turkey.
Recall number
Z-0775-2014
Recall status
Terminated
CDRH record ID
124646
FDA classified
2014-01-16
Recall event number
67065

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗