Skip to content

Public FDA record

Z-0775-2025: Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6), Product Code 35700BAX

Recall record Z-0775-2025 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0775-2025
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Dec 27, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Devices were identified as released after repair without full testing being performed, which includes flow testing.

Source facts

Action
Customers were contacted via telephone beginning 11/19/24. Baxter Healthcare is calling to notify you of a potential issue with the Spectrum infusion pump. One of your devices has been identified as released after repair without full testing being performed, which includes flow testing. Baxter is requesting the return of the impacted infusion pump to perform all tests required and confirm that the infusion pump performs as expected. As part of this process, we need to make you aware of the potential hazard associated with flow related issues. Flow issues may cause excessive therapy, free-flow condition, insufficient therapy or interruption of therapy. These may lead to serious adverse health consequences, including permanent injury and death. Send signed form to Global CORP Service Quality Field Actions Global_CORP_Service_Quality_Field_Actions@baxter.com
Initiated
2024-11-19
Root cause
Process control
Distribution
US Nationwide distribution in the states of AL, IN.
Recall number
Z-0775-2025
Recall status
Open, Classified
CDRH record ID
211363
FDA classified
2024-12-27
Recall event number
95848

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗