Treatment beam may not completely pause while device changes field shape; Potential for mistreatment could occur.
Source facts
Action
On 2/4/05, the firm issued letters to all its direct consignees, informing them of the affected product and providing instructions on the device correction.
Initiated
2005-02-04
Root cause
Software design
Terminated
2006-08-17
Distribution
The firm has indicated that product was released for distribution to 143 direct consignees in US and to 25 foreign countries via international distribution centers.
Foreign distribution include: Australia, Bahrain, Belgium, Canada, Czech Republic, Egypt, Finland, France, Germany, Iceland, India, Israel, Italy, Japan, Kazakhstan, Latvia, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, and United Kingdom.
Recall number
Z-0777-05
Recall status
Terminated
CDRH record ID
37436
FDA classified
2005-04-27
Recall event number
31234