Skip to content

Public FDA record

Z-0777-05: Varian brand Clinac 600C (CD, EX) and Portal Vison with Dynamic MLC or RPM system capability

Recall record Z-0777-05 for Varian Medical Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0777-05
Company in FDA record
Varian Medical Systems Inc
FDA classification date
Apr 27, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Treatment beam may not completely pause while device changes field shape; Potential for mistreatment could occur.

Source facts

Action
On 2/4/05, the firm issued letters to all its direct consignees, informing them of the affected product and providing instructions on the device correction.
Initiated
2005-02-04
Root cause
Software design
Terminated
2006-08-17
Distribution
The firm has indicated that product was released for distribution to 143 direct consignees in US and to 25 foreign countries via international distribution centers. Foreign distribution include: Australia, Bahrain, Belgium, Canada, Czech Republic, Egypt, Finland, France, Germany, Iceland, India, Israel, Italy, Japan, Kazakhstan, Latvia, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, and United Kingdom.
Recall number
Z-0777-05
Recall status
Terminated
CDRH record ID
37436
FDA classified
2005-04-27
Recall event number
31234

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗