Z-0777-2023: BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For in vitro diagnostic use only.

Recall record Z-0777-2023 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0777-2023
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Jan 3, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Beckman Coulter has become aware of an increase in customer complaints generated for intermittent IRISpec CB glucose control false positive results when used in conjunction with iChemVelocity Urine Chemistry Strips PN 800-7212

Source facts

Action
The firm issued a MEDICAL DEVICE RECALL notice to its consignees on 11/18/2022 by mail and email. The notice explained the problem and requested the following actions be taken: "If IRISpec CB controls pass (QC results are within QC acceptance criteria): Continue to process patient samples per IFU (P/N 301-7146BE) If IRISpec CB controls fail (QC results are outside QC acceptance criteria): Refer to Addendum A below for troubleshooting No retrospective review of QC or patient sample results is necessary." "Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter."
Initiated
2022-11-18
Root cause
Under Investigation by firm
Distribution
US and Canada, Brazil, Korea, Mexico, Puerto Rico, Taiwan
Recall number
Z-0777-2023
Recall status
Open, Classified
CDRH record ID
197183
FDA classified
2023-01-03
Recall event number
91175

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗