Z-0778-2024: Accure Laser System, Model: PFMS00004

Recall record Z-0778-2024 for Accure Acne, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0778-2024
Company in FDA record
Accure Acne, Inc
FDA classification date
Jan 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Laser system's spot tracking system may become misaligned resulting in the possible risk of delivery of higher-than-expected laser energy to epidermal tissue, which may cause blistering.

Source facts

Action
On 12/6/23, "Field Safety Alert(s)" were emailed to customers informing them of the following: 1) Following these steps will minimize the likelihood of Region of Interest (ROI) misalignment: Make sure that during treatment that the handpiece tip movement is minimized and to not lift off the tip until the emission is complete and foot pedal released. Remember this order of operations to reduce the potential of misalignment: "Tip down "Pedal down "Emission completes "Pedal up "Tip up 2) Monitor the patient during treatment for unexpected levels of pain. If higher than usual stop treatment and contact the recalling firm. 3) Immediately discontinue therapy if any Skin Overtemperature alarms are observed while treating with the device in Boost mode. This could be an indication that the ROI is misaligned. Contact your Accure representative before any further use of the device. 4) Immediately discontinue therapy if any PETs are observed that are above the green range displayed on the device touchscreen. This could be an indication that the ROI is misaligned. Contact your recalling firm service representative. 5) Ask your patients to contact you immediately if they experience prolonged erythema (longer than 2 hours) or blistering. Blistering is a potential residual risk and is not expected. If it occurs, it may require medical intervention. It could also be an indication that the ROI is misaligned. 6) If ROI misalignment is suspected suspend all further use of the laser and contact the recalling firm. The firm is releasing a software upgrade to maintain the ROI in the intended monitored location regardless of movement of the handpiece tip, and will safely detect and disable the system in the event that the ROI has become misaligned. Acknowledge, and send your response to the firm, see Field Safety Notice. Customers with questions can contact the firm via email at aintintoli@accureacne.com
Initiated
2023-12-06
Root cause
Use error
Distribution
US distribution to states of: CO, CA, CT, NY, NJ, MN, LA, OH, MA, TX
Recall number
Z-0778-2024
Recall status
Open, Classified
CDRH record ID
205151
FDA classified
2024-01-20
Recall event number
93696

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗