Software errors. A number of non-conformances were logged during in-house testing and external validation testing of Brilliance 64, Brilliance Big Bore, and Ingenuity CT systems, equipped with software version 3.5 Beta 2. The non-conformances are due to software errors identified in these device systems.
Source facts
Action
Philips sent an "URGENT - Medical Device Correction" notification letters to all affected customers via Certified mail on October 19, 2011. The letter describes the referenced software issues; provides instructions to the customers as to what they can do in order to address these issues; and informs the customers of the firm's upcoming software field/correction. The letter states that a Philips representative will be contacting all affected customers to install the necessary software upgrade through a Field Change Order. This will be implemented free of charge.
For information or support contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solution Center 1-800-722-9377, option 5.
Initiated
2011-08-29
Root cause
Nonconforming Material/Component
Terminated
2013-02-22
Distribution
Worldwide Distribution - Nationwide (USA) including the states of: IN, MI, MO, NY, OH, and OR and the countries of Germany, Israel and Netherlands.
Recall number
Z-0779-2012
Recall status
Terminated
CDRH record ID
103937
FDA classified
2012-01-17
Recall event number
59887
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.