During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the test screen did not prevent retesting after a failure. Some batteries were therefore retested and passed.
Source facts
Action
Medtronic hand delivered an Urgent Medical Device Recall/Retrieval notice to its sole consignee on 12/07/2023. The notice explained the issue and indicated that the device would be collected and a replacement provided.
Initiated
2023-12-07
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the state of Florida.
Recall number
Z-0779-2024
Recall status
Open, Classified
CDRH record ID
205068
FDA classified
2024-01-22
Recall event number
93663