When the number of fractions is changed for a proton spot beam, the MU are recomputed and rounded if necessary, but the dose is not recalculated as it should be even after saving and re-opening the plan.
Source facts
Action
An Important Safety Notice, LINXIO0017 Proton Sport Beams, Changes in Fractionation not Updating Dose, is in the process of being distributed to affected customers. The release date for this notice is November 25, 2013. The notice included a Confirmation of Receipt which was to be returned to Elekta.
For further questions please call (404) 993-5886.
Initiated
2013-11-25
Root cause
Software Design Change
Terminated
2014-05-13
Distribution
Worldwide Distribution - US Distribution including the states of IL, IN, LA, NJ, TX, WA and WI., and the countries of Belgium, Germany, Italy and Netherlands.
Recall number
Z-0780-2014
Recall status
Terminated
CDRH record ID
124662
FDA classified
2014-01-17
Recall event number
67078