Defective tips have the potential to pipette/dispense volumes 50-100% lower than is specified for the tips.
Source facts
Action
PerkinElmer initiated the recall on Sept 1. 2016 by letter dated . Customers were told of the defect, potential for inaccurate measurements, and requested to inspect for the lots stop using and destroy. Replacement product to be provided. Accounts asked to compelte Response form. Questions call your PerkinElmer representative.
Initiated
2016-10-13
Root cause
Process control
Terminated
2018-09-13
Distribution
IA, MD, NY, WA
Foreign: France, Germany, The Netherlands, Hungary, Singapore
Recall number
Z-0780-2017
Recall status
Terminated
CDRH record ID
150859
FDA record created
2016-12-16
Recall event number
75588