Skip to content

Public FDA record

Z-0780-2020: Stryker T2 Alpha Delta Strike Plate IMN Instruments - part of the IMN Instruments System and is used to s

Recall record Z-0780-2020 for Stryker GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0780-2020
Company in FDA record
Stryker GmbH
FDA record creation date
Jan 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Instrument can break at the level of the thread when being exposed to high forces during nail implantation or removal which can lead to complications during nail insertion and removal.

Source facts

Action
Stryker issued Urgent Medical Device Recall Letter (PFA# 2219868) to US consignees on 5-Dec-2019 , letter states reason for recall, health risk and action to take: 1.Please inform users of this Urgent Medical Device Recall and forward this notice to all those individuals who need to be aware within your organization. 2.Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. A response is required, even though you may not have any physical inventory on site anymore. 3.Quarantine and discontinue use of the recalled devices. 4. Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Recall Business Reply. An OUS customer letter was also issued at the country level.
Initiated
2019-12-05
Root cause
Under Investigation by firm
Terminated
2020-10-09
Distribution
Worldwide distribution - US Nationwide distribution and countries of Switzerland, France, Spain, United Kingdom, Italy, Sweden, Netherlands, Australia.
Recall number
Z-0780-2020
Recall status
Terminated
CDRH record ID
178376
FDA record created
2020-01-07
Recall event number
84458

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗