Medtronic has discovered an instance in which the Intellis A710 Clinician Programmer Application (CP App) could no longer establish a session with the Implantable Neurostimulator (INS) which resulted in the inability for the clinician to adjust stimulation parameters. In the event of this issue, the patient may receive inadequate or excessive therapy and experience non-optimal pain relief.
Source facts
Action
OUS communication to a single German consignee will be initiated via regionally approved methods beginning 19-Dec-2023. A field service technician will go to the site and reset the INS to resolve the issue.
Initiated
2023-12-19
Root cause
Under Investigation by firm
Distribution
Germany
Recall number
Z-0780-2024
Recall status
Open, Classified
CDRH record ID
205296