Z-0781-2021: Thyroid Pack CKHA-LF, Model DYNJ50352B - Product Usage: This is a type of convenience kit used in various

Recall record Z-0781-2021 for Medline Industries Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0781-2021
Company in FDA record
Medline Industries Inc
FDA record creation date
Jan 7, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The cylindrical sponge component is not x-ray detectable.

Source facts

Action
Customers were notified by email on 11/13/2020. Customers are asked to return the response form via fax or email. The recalling firm will send subsequent letters to consignees that have not provided a response to the recall notification within the allotted time. No product is to be returned to Medline Industries. Consignees are instructed to apply labels to their affected kits and return to inventory. The label indicates that there is a recalled component within the pack, not to use it, and replace from a different sterile supply. No other components within the kit are affected.
Initiated
2020-11-13
Root cause
Process control
Terminated
2021-10-26
Distribution
Foreign distribution only - Canada.
Recall number
Z-0781-2021
Recall status
Terminated
CDRH record ID
184623
FDA record created
2021-01-07
Recall event number
86826

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗