Potential separation of the port base from the port cover after implantation.
Source facts
Action
Letters sent 3/11/2005 via Federal Express to hospitals receiving affected product. Customers instructed to remove/return product, notify physicians, evaluate patients and explant devices whenever possible.
Initiated
2005-02-25
Root cause
Other
Terminated
2006-08-09
Distribution
1215 units shipped to 123 hospitals throughout the U.S. 2 units shipped to Boston Scientific in South Africa where 1 unit had been used as a demo (non-human use) and 1 units is in BSC control
Recall number
Z-0782-05
Recall status
Terminated
CDRH record ID
37535
FDA classified
2005-04-28
Recall event number
31251