Z-0782-2024: PANTHER FUSION System REF 9121010000 The Panther System is an integrated nucleic acid testing system wh

Recall record Z-0782-2024 for Hologic, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0782-2024
Company in FDA record
Hologic, Inc.
FDA classification date
Jan 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to potential false positive Flu B results when a sample was also SARS-CoV-2 positive.

Source facts

Action
On 12/15/2023, the firm send an "URGENT: MEDICAL DEVICE RECALL" Letter via FedEx 2-day delivery to customer informing them that the firm has became aware of potential erroneous influenza B virus (FLU B) positive results when testing with the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay on the Panther Fusion System. Customers are instructed to: 1. They may continue to test with the Panther Fusion SARS-CoV-2/Flu A/B/RTSV assay on the Panther Fusion System. 2. Hologic has partnered with Novasyte, an IQVIA MedTech company, to assist in the recall. Acknowledge receipt of the notification by completing the Customer Acknowledgement Form located at https;//iqvia-response.my.site.com/mt/fca?cid=HPAN23 3. To ensure visibility of the issue, Hologic recommends posting the notification on or near the Panther Fusion Systems that are running the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay. 4. If customers suspect a false Flu B positive result in combination with SARS-CoV-2 positive result (dual positive), contact Hologic Technical Support for evaluation and/or troubleshooting. For questions or assistance, contact Hologic Technical Support. Country-specific telephone and email contact information for Technical Support can be found at www.hologic.com/support.
Initiated
2023-12-15
Root cause
Under Investigation by firm
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AR, CA, CO, CT, FL, IL, IN, MA, ME, MI, MN, MO, MT, NC, ND, NH, NJ, NM, NV, NY, OH, PA, Puerto Rico, TX, UT, VA, VI, WA, WI, and WV. The countries of Austria, Australia, Belgium, Dominican Republic, Switzerland, Germany, Denmark, Spain, France, Great Britain, Hong Kong, Ireland, Italy, Korea, Kuwait, The Netherlands, Norway, New Zealand, Portugal, Qatar, and Sweden.
Recall number
Z-0782-2024
Recall status
Open, Classified
CDRH record ID
204926
FDA classified
2024-01-23
Recall event number
93593

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗