Open original FDA record ↗Recall record
- FDA identifier
- Z-0782-2024
- Company in FDA record
- Hologic, Inc.
- FDA classification date
- Jan 23, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to potential false positive Flu B results when a sample was also SARS-CoV-2 positive.
Source facts
- Action
- On 12/15/2023, the firm send an "URGENT: MEDICAL DEVICE RECALL" Letter via FedEx 2-day delivery to customer informing them that the firm has became aware of potential erroneous influenza B virus (FLU B) positive results when testing with the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay on the Panther Fusion System. Customers are instructed to: 1. They may continue to test with the Panther Fusion SARS-CoV-2/Flu A/B/RTSV assay on the Panther Fusion System. 2. Hologic has partnered with Novasyte, an IQVIA MedTech company, to assist in the recall. Acknowledge receipt of the notification by completing the Customer Acknowledgement Form located at https;//iqvia-response.my.site.com/mt/fca?cid=HPAN23 3. To ensure visibility of the issue, Hologic recommends posting the notification on or near the Panther Fusion Systems that are running the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay. 4. If customers suspect a false Flu B positive result in combination with SARS-CoV-2 positive result (dual positive), contact Hologic Technical Support for evaluation and/or troubleshooting. For questions or assistance, contact Hologic Technical Support. Country-specific telephone and email contact information for Technical Support can be found at www.hologic.com/support.
- Initiated
- 2023-12-15
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AR, CA, CO, CT, FL, IL, IN, MA, ME, MI, MN, MO, MT, NC, ND, NH, NJ, NM, NV, NY, OH, PA, Puerto Rico, TX, UT, VA, VI, WA, WI, and WV. The countries of Austria, Australia, Belgium, Dominican Republic, Switzerland, Germany, Denmark, Spain, France, Great Britain, Hong Kong, Ireland, Italy, Korea, Kuwait, The Netherlands, Norway, New Zealand, Portugal, Qatar, and Sweden.
- Recall number
- Z-0782-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204926
- FDA classified
- 2024-01-23
- Recall event number
- 93593