Open original FDA record ↗Recall record
- FDA identifier
- Z-0783-2013
- Company in FDA record
- The Anspach Effort, Inc.
- FDA classification date
- Feb 4, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Contact between the bur and bur guard could generate metal fragments that may or may not be visible to the surgeon and can potentially remain in the surgical site. The materials used to fabricate the bur guard are not traceable, design validation did not effectively evaluate adequate protection of adjacent tissue and inspection results for each of these bur guards were not documented.
Source facts
- Action
- The firm, Anspach Effort, Inc. sent an "URGENT: Medical Device Removal" letter dated November 7, 2012, to its customer. The letter identified the product, problem and actions to be taken. The customer was instructed to do the following: 1) Screen their inventory and remove and return all products immediately. 2) Complete and return the attached Customer Reply Form via fax or email to the address provided on the form. Should the customer have any questions, please contact Anspach Product Support at (800) 327-6887.
- Initiated
- 2012-09-05
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-08-13
- Distribution
- Distributed in the state of Massachusetts
- Recall number
- Z-0783-2013
- Recall status
- Terminated
- CDRH record ID
- 114783
- FDA classified
- 2013-02-04
- Recall event number
- 63793