Z-0783-2014: Zimmer Segmental Systems; Polyethylene Inserts, Size B. Designed to address significant bone loss result

Recall record Z-0783-2014 for Zimmer, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0783-2014
Company in FDA record
Zimmer, Inc.
FDA classification date
Jan 17, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may place excessive loading on the polyethylene insert.

Source facts

Action
Depth of recall " Notifications were sent to customers, surgeons and risk managers, which use the affected product. Distributors were also informed of the notifications sent to surgeons and risk managers. The customers were identified by performing an invoice search and distribution history for the Polyethylene Inserts (00-5850-012-95 and 00-5850-013-95) since product launch. Recall Communication " All distributors were notified via electronic mail. Distributors that have received affected inventory were also notified via courier. Hospital risk managers and surgeons were also notified via courier. On 12/19/13, hospital risk managers and surgeons were provided with a notification identifying the issue and their responsibilities. These responsibilities include ensuring the affected personnel are aware of the contents of the notification. On 12/19/13, Distributors were sent a notification identifying the issue and to provide a copy of the notification sent to surgeons and risk managers. The distributors will also be instructed as needed to assist with identifying a comprehensive list of surgeons. " Communications outside of the United States will occur approximately two weeks after the United States communications. Effectiveness Checks Zimmer will conduct effectiveness checks in the following way: "Surgeons/Hospitals o100% of the FedEx notifications will be accounted for. oSecond and third notification attempts will be made. oFailed delivery attempts following the third attempt will result in the account being deemed as nonresponsive.
Initiated
2013-12-19
Root cause
Device Design
Terminated
2015-02-24
Distribution
Worldwide Distribution. USA nationwide and Germany, with further distribution into EU countries.
Recall number
Z-0783-2014
Recall status
Terminated
CDRH record ID
123738
FDA classified
2014-01-17
Recall event number
66846

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗