Z-0783-2022: (1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrie

Recall record Z-0783-2022 for Hill-Rom, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0783-2022
Company in FDA record
Hill-Rom, Inc.
FDA record creation date
Mar 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately and could loosen over time causing the rail system to fall, potentially causing injury to caregivers and/or patients

Source facts

Action
HillRom issued Urgent Medical Device Correction letter(FA-2021-11-004-LUL-002) on the 14th of January via UPS Ground service. Letter states reason for recall, health risk and action to take: Actions to be taken by Customer: Please identify if you have the potentially affected product in your facility (See affected device listing above). Do not use your traverse carrier until inspected by a Hillrom Technician. Return the response form below within two weeks. Upon receipt of the response form, Hillrom will contact you to schedule a visit to inspect your carriers. Should incorrect, non-conforming screws be present, these will be exchanged by an approved Hillrom technician. Action to be taken by the Distributor: Please share this Field Safety Notice with your end users and complete the attached response form and return to hillromLUL002@sedgwick.com within two weeks. Hillrom will provide replacement screws for Distributors to facilitate exchange among your end users. Action to be taken by Hillrom: Hillrom will arrange a site visit to inspect the potentially affected units. If short screws are identified, they will be exchanged with those of correct length. Contact reference person If you have any questions regarding this safety notice, please contact Hillrom Technical Support at 800-445-3720 or technical.support@hillrom.com .
Initiated
2022-01-14
Root cause
Nonconforming Material/Component
Terminated
2024-11-07
Distribution
Worldwide distribution - US Nationwide and the countries of AT, CA, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IT, JP, NL, NO, NZ, RU, SE, SG, SK.
Recall number
Z-0783-2022
Recall status
Terminated
CDRH record ID
191947
FDA record created
2022-03-11
Recall event number
89562

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗